150 Keysight N9038A MXE Service Guide
Input Selection & Level Control
Input Selection & Level Control Section Troubleshooting
Figure 4-16
A23 Limiter Assembly Connections
4.
, disconnect W6 from the A23 Limiter assembly
LED Drive Output (P2).
5.
Turn the instrument on and allow it to complete its boot up process,
ignoring any error messages.
6.
Connect a signal generator to the instrument RF Input 2, as shown in
.
7.
Tune the signal generator to 100 MHz with an amplitude of
−
20 dBm.
8.
Connect a spectrum analyzer to the A23 Limiter assembly Output (J103),
as shown in
.
9.
Tune the spectrum analyzer to a frequency of 100 MHz with a span of
10 MHz.
10.
Verify that the signal level on the spectrum analyzer is at
−
29 dBm
(±1.5 dB), as shown in
, allowing for any additional test cable
loss.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...