Instrument Messages
Event Messages
–221
Settings conflict;
Signal Track is
turned off when Zero
Span is selected
Signal Track is not available when you have selected Zero Span. So if Zero
Span is entered while in Signal Track is On, Signal Track is turned off
–221
Settings Conflict;
Span limited to XXX
–221
Settings conflict;
Span Zoom is not
available in Zero
Span
Span Zoom does not work with a time domain X-axis. You must select a span
greater then 0 Hz
–221
Settings conflict;
Step keys are not
available to modify
this function
You should select a specific value for this function. Using the
Up/Down
step keys to scroll through values is not allowed
–221
Settings conflict;
Sweep Setup only
available in swept
measurements
The current measurement uses FFT mode, so does not use the Sweep Setup
menu
–221
Settings conflict;
Sweep Time cannot be
auto-coupled in FFT
sweeps
The sweep time for FFT sweeps is set by the calculations. So sweep time
settings cannot be adjusted
–221
Settings conflict;
Sweep Time cannot be
auto-coupled while in
Zero Span
You cannot send the remote command to set the sweep time to auto while
you are in zero span
–221
Settings conflict;
Sweep Time cannot be
set while in FFT
sweeps
The sweep time for FFT sweeps is set by the calculations. It cannot be
manually controlled
–221
Settings conflict;
Swept IF Gain High
not available when
FFT IF Gain = Manual
Low
When FFT IF Gain is manually set to Low, you cannot set the Swept IF Gain to
High because that would make the Reference Level couplings wrong in
swept mode
–221
Settings conflict;
Swept LO not
available when freq
mode is Fixed
SCPI only message. The LO Mode cannot be set to Swept when the freq
mode is set to fixed. Change the freq mode away from fixed, or perform the
measurement at several fixed frequencies
–221
Settings conflict;
Swept Type=Swept is
not available while
in Gated FFT
If you have selected gated FFT then you are using the FFT sweep type and
you cannot select the swept type of sweeping
Err#
Message
Verbose/Correction Information
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...