716 Keysight N9038A MXE Service Guide
Tests
Quasi-Peak Detector Accuracy
27.
Setup the IGUU 2916 EMI pulse generator as follows:
a.
Aux. Generator:
Band B
b.
Pulse Frequency:
100 Hz
c.
Amplitude:
40 dB
μ
V
28.
Move the cable on the IGUU 2916 EMI pulse generator to Output Auxiliary
Generator Band A/B/C/D.
29.
Change the EMI receiver input attenuation to 10 dB by pressing
AMPTD Y
Scale
,
Attenuation
,
10 dB
.
30.
Take a single sweep on the EMI receiver by pressing
Restart
.
31.
Reset the delta marker by pressing
Marker
,
Delta
.
32.
Change the Pulse Frequency on the IGUU 2916 EMI pulse generator to
1000 Hz
.
33.
Take a single sweep on the EMI receiver by pressing
Restart
.
34.
Once the sweep on the EMI receiver has completed and the auto peak
value is found record the marker delta amplitude value for the pulse
frequency of 1000 Hz in
Band C Testing
1.
Set the IGUU 2916 EMI pulse generator to output a 100 MHz Sine Wave
signal.
2.
Connect the IGUU 2916 EMI pulse generator Output Sine Wave Generator
to the EMI receiver RF Input.
3.
Preset the EMI receiver by pressing
Mode
Preset
.
4.
Tune the receiver to 100 MHz with a span of 0 Hz by pressing
FREQ
Channel
,
Center Freq
,
100 MHz
then
SPAN X Scale
,
Zero Span
.
5.
Set the Sweeptime to 2 seconds by pressing
Sweep / Control
,
2 s
.
6.
Set the amplitude units to dB
μ
V by pressing
AMPTD Y Scale
,
More
,
Y Axis
Unit
,
More
,
dB
μ
V
.
7.
Set the reference level to 80 dB
μ
V by pressing
AMPTD Y Scale
,
80 dB
μ
V
.
8.
Change the detector used by the EMI receiver to Quasi-Peak by pressing
Trace / Detector
,
More
,
Detector
,
More
,
Quasi Peak
.
9.
Turn on the marker by pressing
Marker
.
10.
Record the marker value in
as the Measured value for Band C.
11.
Calculate the Error with the following formula and enter the value in
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...