666 Keysight N9038A MXE Service Guide
Tests
Functional Test Versus Performance Verification
Functional Test Versus Performance Verification
Functional tests use a minimum set of test equipment to check a much smaller
range of parameters (and a limited number of data points for each parameter)
than do performance verification tests. Functional tests use limits that are
wider than the published specifications; measurement uncertainty analysis is
not
available for functional tests.
Performance verification tests span a wide range of instrument parameters and
provide the highest level of confidence that the instrument conforms to
published specifications. These tests can be time consuming and require
extensive test equipment.
If a functional test does not pass, you must run performance verification tests to determine
whether a problem exists.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...