Keysight N9038A MXE Service Guide 291
Front End Control
A15 Front End Control Assembly Troubleshooting
A12 YTF Preselector Alignment
The alignment of the A12 YTF Preselector is performed by the use of an internal
instrument routine and requires no additional test equipment. Use the
following procedure to perform the A12 YTF Preselector alignment routine:
1.
If the Front End Control Test board was being used just prior to the running
of this routine turn the instrument off, disconnect the Front End Control
Test board from the instrument, and plug all of the instrument internal
cables back into the A15 Front End Control assembly.
2.
Then turn the instrument on and allow it to complete its boot up process.
3.
Run the A12 YTF Preselector alignment routine by pressing
System
,
Alignments
,
More
,
Advanced
,
Characterize Preselector
.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...