710 Keysight N9038A MXE Service Guide
Tests
Quasi-Peak Detector Accuracy
Figure 22-10
Typical Equipment Setup
Sine Wave Generator Characterization
1.
Zero and calibrate the power meter.
2.
Connect the power meter to the Output Sine Wave Generator of the IGUU
2916 EMI pulse generator as shown in
.
Figure 22-11
Sine Wave Generator Characterization
3.
Set the IGUU 2916 EMI pulse generator to output a 100 kHz Sine Wave
signal.
4.
After allowing the power meter to settle, record the reading in
under the Power Meter column.
5.
Set the IGUU 2916 EMI pulse generator to output a 10 MHz Sine Wave
signal.
6.
After allowing the power meter to settle, record the reading in
under the Power Meter column.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...