184 Keysight N9038A MXE Service Guide
RF Preselector Section
RF Preselector Section Troubleshooting
Output Signal Routing and Preamplifiers
There are two main signal paths through the output routing of the A21 RF
Preselector Input assembly, and a preamplifier for each, to be verified, as seen
Figure 5-11
A21 RF Preselector Input Assembly Output Switching
Verifying Output Signal Routing and Preamplifier Gain
1.
Remove the instrument dress cover (MP24) and the top brace (MP10).
Refer to
Chapter 18, “Assembly Replacement Procedures”
for instructions
on removing these covers.
2.
, remove W33, W35, and W36 from
the A21 RF Preselector Input assembly.
3.
Turn the instrument on and allow it to complete its boot up process,
ignoring any error messages.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...