Keysight N9038A MXE Service Guide 117
Instrument Messages
Condition Messages
For example, error 207 indicates a Burst Not Found condition has been
detected, error 1207 indicates that failure has been cleared.
This register is summarized as bit 0 of the
STATus:QUEStionable:INTegrity
register, as described in the section
“Condition Errors 101 to 199, Measurement Integrity” on page 113
Err#
Bit in
status
register
Message
Error or
Warning
More Information
203
0
unused E
205
1
unused
E
207
2
Burst Not Found
E
The burst signal cannot be detected because of
inappropriate parameter settings or incorrect signal.
An in appropriate parameter setting could cause the
signal to be partially, rather than fully, on the display,
Burst Search Threshold and/or Burst Search Length
may need to be adjusted.
An incorrect signal could have either insufficient power,
the rising or falling edges cannot be detected, or the
burst is less than 126
microseconds.
Carrier signal is not actually bursted.
W-CDMA: Either the signal being analyzed has
insufficient power, the rising or falling edges cannot be
detected, or the burst is less than 126
microseconds.
W-CDMA: Cannot synchronize measurement with
PRACH channel for Power Control measurement,
because the signal cannot be found. Make sure PRACH
is present in the W-CDMA uplink signal, and that the
preamble signature and scramble code are set
correctly.
GSM: Data was acquired but a GSM burst was not
found, with the timeslot mode disabled.
NADC, PDC: A valid burst is not found when the Device
is MS.
1xEV-DO: Data was acquired but a 1xEV burst was not
found, with the timeslot mode disabled.
Bluetooth
: The burst that has been found does not
correspond to the currently selected
Bluetooth
packet
type (the burst length may be too short).
WLAN: The instrument cannot find a valid WLAN burst.
You may need to extend the search length.
In PSA, this error was reported as one of the following
error numbers: 10772, 13104, 10160, 10286, 10420,
10454, 10614, 10904, 10928, 13074, 10287
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...