126 Keysight N9038A MXE Service Guide
Instrument Messages
Condition Messages
Condition Errors 601 to 699, Error Summaries
This series of errors corresponds to the bits in the
STATus:QUEStionable
register, read with a
STATus:QUEStionable?
event query or a
STATus:QUEStionable:CONDition?
query.
The second column in the table shows the corresponding bit in the status
register. These bits do not have any corresponding error messages, although
error numbers have been reserved for them as seen in the Err# column; they
are status bits only.
The bits in the
STATus:QUEStionable
register are “summary bits” for
registers at a lower level. Note that these summary bits summarize the state
and history of the event registers at the lower level. This is true even for bits in
the
STATus:QUEStionable
condition register. This means that:
— The summary bits read by the
STATus:QUEStionable: CONDition?
query are true if any event bits are set in any of the :QUEStionable
sub-registers :POWer, :TEMPerature, FREQuency, CALibration or :INTegrity.
— The summary bits read by the
STATus:QUEStionable?
event query are
true if any event bit has undergone a false-to-true transition with the
PTRansition filter set, or a true-to-false transition with the NTRansition filter
set, in any of the :QUEStionable sub-registers :POWer, :TEMPerature,
FREQuency, CALibration or :INTegrity.
Thus, the summary bits
cannot
be used to determine the current state of a
lower level condition bit; only the state and history of the lower level event bits.
523
11
unused
525
12
unused
527
13
unused
529
14
unused
Err#
Bit in
status
register
Message
Error or
Warning
More Information
Err#
Bit in
status
register
Message
Error or
Warning
More Information
601
0
unused
603
1
unused
605
2
unused
607
3
Power
status bit
only
This bit is the summary bit for the
STATus:QUEStionable:POWer sub-register.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...