10-30
SCPI Command Reference
Models 2510 and 2510-AT User’s Manual
:RANGe <n>
[:SENSe[1]]:TEMPerature:RTD:RANGe <n>
Set RTD resistance range
Parameters
<n> =
0 to 1000
Set RTD resistance range
(100 or 1000)
MINimum
0
MAXimum
1000
DEFault
100
Query
:RANGe?
Query sensor resistance range
Description
This command programs the temperature sensor resistance range. It
affects only RTD sensor types. Nominal ranges are: 100 and 1000.
:ALPHa <n>
[:SENSe[1]]:TEMPerature:RTD:ALPHa <n>
Program RTD sensor alpha
Parameters
<n> =
0 to 0.01
RTD
α
coefficient
MINimum
0
MAXimum
0.01
DEFault
0.00385
Query
:ALPHa?
Query RTD
α
coefficient
Description
This command programs the RTD temperature sensor
α
coefficient for
the USER type RTD sensor.
NOTE
See
“Sensor coefficients,” page 4-14
:BETA <n>
[:SENSe[1]]:TEMPerature:RTD:BETA <n>
Program RTD sensor beta
Parameters
<n> =
0 to 1
RTD
β
coefficient
MINimum
0
MAXimum
1
DEFault
0.111
Query
:BETA?
Query RTD
β
coefficient
Description
This command programs the RTD temperature sensor
β
coefficient for
the USER type RTD sensor.