Models 2510 and 2510-AT User’s Manual
SCPI Command Reference
10-41
Voltage function
[:AMPLitude] <n>
:SOURce[1]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <n> Program voltage source value
Parameters
<n> =
-10 to +10V
Set voltage source value
MINimum
-10
MAXimum
+10
DEFault
0.0
Query
[:AMPLitude]?
Query voltage source value
Description
This command programs the amplitude (setpoint) for the voltage source.
The voltage amplitude can be set over a range of ±10V, but the value
cannot exceed the voltage protection limit.
:LCONstants
:SOURce[1]:VOLTage:LCONstants[:GAIN] <n>
Program proportional constant
:SOURce[1]:VOLTage:LCONstants:DERvative <n>
Program derivative constant
:SOURce[1]:VOLTage:LCONstants:INTegral <n>
Program integral constant
Parameters
[:GAIN] <n> =
0 to 1e5
Set proportional constant value
MINimum
0
MAXimum
1e5
DEFault
10.0
:DERivative <n> = 0 to 1e5
Set derivative constant value
MINimum
0
MAXimum
1e5
DEFault
0.0
:INTegral <n> =
0 to 1e5
Set integral constant value
MINimum
0
MAXimum
1e5
DEFault
20.0
Query
[:GAIN]?
Query proportional constant
:DERivative?
Query derivative constant
:INTegral?
Query integral constant
Description
These commands program the voltage function PID loop constants,
which must be set to appropriate values for suitable control characteris-
tics. See
appropriate values.