10-42
SCPI Command Reference
Models 2510 and 2510-AT User’s Manual
:PROTection
:SOURce[1]:VOLTage:PROTection[:LEVel] <n>
Program voltage limit
:SOURce[1]:VOLTage:PROTection:TRIPped?
Query if voltage limit exceeded
Parameters
<n> =
0.5 to +10.5V
Set voltage limit value
MINimum
0.5
MAXimum
+10.5
DEFault
+10.5
Query
[:LEVel[?
Query voltage limit
:TRIPped?
Query if voltage limit exceeded (1 =
exceeded)
Description
The :LEVel command programs the voltage limit for the voltage func-
tion. The :TRIPPed? query returns a value of 1 if the voltage limit is
exceeded or a value of 0 if it is not exceeded. These protection com-
mands are active only for the voltage function and are absolute. As a
result, there are no high/low limits or :STATe commands since voltage
protection is always active.
Setpoint tolerance
[:PERCent] <n>
:SOURce[1]:STOLerance[:PERCent] <n>
Program tolerance window
Parameters
<n> =
0 to 100%
Program tolerance (% of range)
MINimum
0
MAXimum
100
DEFault
0.5
Query
[:PERCent]?
Query percent tolerance.
Description
This command is used to program the percentage of range for the set-
point tolerance. Ranges depend on selected function as follows:
Function
Range
Temperature
275°C
275K
495°F
Voltage
20V
Current
10A
Resistance
1M
Ω
(100k
Ω
thermistor)
100k
Ω
(10k
Ω
thermistor)
10k
Ω
(1k
Ω
thermistor or RTD)
1k
Ω
(100
Ω
thermistor or RTD)
Summary of Contents for TEC SourceMeter 2510
Page 194: ...A Specifications...
Page 197: ...B StatusandErrorMessages...
Page 206: ...C IEEE 488BusOverview...
Page 221: ...D IEEE 488andSCPI ConformanceInformation...
Page 225: ...E ExamplePrograms...
Page 233: ...F GPIB488 1Protocol...