1-18
Getting Started
Models 2510 and 2510-AT User’s Manual
Setting
Front panel BENCH and GPIB default*
Current function
Setpoint
0A
Current protection limit
2A
Proportional constant
10
Integral constant
20
Derivative constant
0
Resistance function
Setpoint
100
Ω
(100
Ω
RTD or thermistor)
1k
Ω
(1k
Ω
RTD or thermistor)
10k
Ω
(10k
Ω
thermistor)
100k
Ω
(100k
Ω
thermistor)
Protection high limit
200
Ω
(100
Ω
RTD or thermistor)
2k
Ω
(1k
Ω
RTD or thermistor)
20k
Ω
(10k
Ω
thermistor)
200k
Ω
(100k
Ω
thermistor)
Protection low limit
50
Ω
(100
Ω
RTD or thermistor)
500
Ω
(1k
Ω
RTD or thermistor)
5k
Ω
(10k
Ω
thermistor),
50k
Ω
(100k
Ω
thermistor)
Protection state
Enabled
Sensor current
2.5mA (100
Ω
range )
833.3µA (1k
Ω
range)
100µA (10k
Ω
range)
33.3µA (100k
Ω
range)
Sensor current auto state
On
Sensor type
Thermistor
Sense mode
2-wire
Thermistor range
10k
Ω
Proportional constant
20
Integral constant
0.6
Derivative constant
0
Output On/Off
Off
Digital I/O enable line
Off
Peltier (cooler) ground connect mode
Disabled
Digital I/O port output value
15
Display format**
Engineering units
Setpoint tolerance percentage
0.5%
Window size
5
*
* BENCH defaults are the same as :SYSTem:PRESet defaults (
). GPIB defaults are the same
as *RST defaults (
** Not affected by BENCH or GPIB RESET.
Table 1-3 (continued)
Factory front panel default settings
Summary of Contents for TEC SourceMeter 2510