7-4
Status Structure
Models 2510 and 2510-AT User’s Manual
Clearing registers and queues
When the Model 2510 is turned on, the bits of all registers in the status structure are
cleared (reset to 0), and the two queues are empty. Commands to reset the event and event
enable registers, and the Error Queue are listed in
. In addition to these com-
mands, any enable register can be reset by sending the 0 parameter value with the individ-
ual command to program the register.
NOTE
SYSTem:PRESet and *RST have no effect on status structure registers and
queues.
Table 7-1
Common and SCPI commands to reset registers and clear queues
Commands
Description
Ref.
To Reset Registers:
*CLS
Reset all bits of the following event registers to 0:
Note 1
Standard Event Register
Operation Event Register
Measurement Event Register
Questionable Event Register
:STATus:PRESet
Reset all bits of the following enable registers to 0:
Operation Event Enable Register
Measurement Event Enable Register
Questionable Event Enable Register
Note 1
To Clear Error Queue:
*CLS
Clear all messages from Error Queue
Note 2
:STATus:QUEue:CLEar
Clear messages from Error Queue
Note 3
:SYSTem:ERRor:CLEar
Clear messages from Error Queue
Note 3
Notes:
1. The Standard Event Enable Register is not reset by STATus:PRESet or *CLS. Send the 0 parameter value
with *ESE to reset all bits of that enable register to 0 (see
“Status byte and service request (SRQ),”
2. STATus:PRESet has no effect on the Error Queue.
3. Use either of the two clear commands to clear the Error Queue.