Models 2510 and 2510-AT User’s Manual
SCPI Command Reference
10-39
:LCONstants
:SOURce[1]:TEMPerature:LCONstants[:GAIN] <n>
Program proportional constant
:SOURce[1]:TEMPerature:LCONstants:DERvative <n>
Program derivative constant
:SOURce[1]:TEMPerature:LCONstants:INTegral <n>
Program integral constant
Parameters
[:GAIN] <n> =
0 to 1e5
Set proportional constant value
MINimum
0
MAXimum
1e5
DEFault
20.0
:DERivative <n> = 0 to 1e5
Set derivative constant value
MINimum
0
MAXimum
1e5
DEFault
0.0
:INTegral <n> =
0 to 1e5
Set integral constant value
MINimum
0
MAXimum
1e5
DEFault
0.6
Query
[:GAIN]?
Query proportional constant
:DERivative?
Query derivative constant
:INTegral?
Query integral constant
Description
These commands program the temperature function PID loop constants,
which must be set to appropriate values for suitable control characteris-
tics. See
appropriate values.