Models 2510 and 2510-AT User’s Manual
PID Control Concepts
4-7
Figure 4-4
Response comparison example 1 (short Lag and Tau times)
Laser Diode Module with Minimum Settling Time
23.0
22.0
0
5
10
15
20
25
24.0
25.0
26.0
30
22.5
23.5
24.5
25.5
26.5
A. For 3
°
C step: Settled to
±
1.0% (
±
0.030
°
C) in 8.54 seconds
Settled to
±
0.1% (
±
0.003
°
C) in 11.14 seconds
Laser Diode Module with Minimum Overshoot
23.0
22.0
0
5
10
15
20
25
24.0
25.0
26.0
30
22.5
23.5
24.5
25.5
26.5
B. For a 3
°
C step: Settled to
±
1.0% (
±
0.030
°
C) in 15.32 seconds
Settled to
±
0.1% (
±
0.003
°
C) in 27.32 seconds
Settled to 1.0%
(0.030
°
C)
Settled to 0.1%
(0.003
°
C)
Settled to 1.0%
(0.030
°
C)
Settled to 0.1%
(0.003
°
C)
Summary of Contents for TEC SourceMeter 2510
Page 194: ...A Specifications...
Page 197: ...B StatusandErrorMessages...
Page 206: ...C IEEE 488BusOverview...
Page 221: ...D IEEE 488andSCPI ConformanceInformation...
Page 225: ...E ExamplePrograms...
Page 233: ...F GPIB488 1Protocol...