:SOURce[1]
:TEMPerature
:PROTection
Set temperature protection.
✓
[:HIGH]
Upper temperature limit.
✓
[:LEVel] <n>
Set upper temperature limit.
<n> = -50 to +250°C, 223.15 to 523.15K,
-58 to 482°F | MIN | MAX | DEF
50.0°C,
323.15K,
122°F
✓
[:LEVel]?
Query upper protection temperature limit.
✓
:TRIPped?
Query if upper limit is exceeded. (1 = yes;
0 = no).
✓
:LOW
Lower temperature limit.
0.0
✓
[:LEVel] <n>
Set lower temperature limit.
<n> = -50 to +250°C, 223.15 to 523.15K,
-58 to 482°F | MIN | MAX | DEF
0°C,
273.15K,
32°F
✓
[:LEVel]?
Query lower protection temperature limit.
✓
:TRIPped?
Query if lower limit exceeded. (1 = yes;
0 = no).
✓
:STATe <b>
Enable/disable temperature protection.
(ON | OFF)
ON
:STATe?
Query temperature protection state.
[:SPOint] <n>
Program temperature setpoint.
<n> = -50 to +225°C, 223.15 to 498.15K,
-58 to 437°F | MIN | MAX | DEF
25°C,
298.15K,
77°F
✓
[:SPOint]?
Query temperature setpoint.
✓