Series 2280 Precision DC Power Supply Reference Manual
Section 7: SCPI command reference
077085503 / March 2019
7-91
Detail
Save the specified list to the USB flash drive.
Make sure that the external USB flash drive is connected to the instrument before you send this
command. An unstable connection between the USB flash drive and the instrument will cause an
error 520: "No USB flash drive found."
Example
:LIST:SAVE:USB 1, "mylist"
Export the list 1 to the external USB flash drive as
mylist
.
Also see
(on page 7-85)
[:SOURce[n]]:LIST:STATe
This command enables or disables the list mode.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 (OFF)
Usage
[:SOURce[n]]:LIST:STATe <value>
[:SOURce[n]]:LIST:STATe?
[n]
Channel number: For single-channel instruments, the value is
1
<value>
Enable the list mode:
OFF
or
0
Disable the list mode:
ON
or
1
Details
This command enables or disables the list mode.
The list only works in the concurrent function. If the function is not concurrent, sending the
[:SOURce[n]]:LIST:STATe
command will result in the error 209, "Change Sense Function to V+I
(CONCurrent) before running a list."