Series 2280 Precision DC Power Supply Reference Manual
Section 7: SCPI command reference
077085503 / March 2019
7-69
:SENSe[n]:<function>:RANGe:AUTO
This command determines if the range for measurements of the selected function is selected manually or
automatically.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Saved setup
1 (ON)
Usage
:SENSe[n]:<function>:RANGe:AUTO <state>
:SENSe[n]:<function>:RANGe:AUTO?
[n]
Channel number: For single-channel instruments, the value is
1
<function>
The measurement function to which this setting applies:
Concurrent:
CONCurrent[:DC]
Current:
CURRent[:DC]
<state>
Set the measurement range manually:
OFF
or
0
Set the measurement range automatically:
ON
or
1
Details
This command determines how the range is selected.
When this command is set to off, you must set the range. If you do not set the range, the instrument
remains at the range that was last selected by autorange.
When this command is set to on, the instrument automatically goes to the most sensitive range to
perform the measurement.
If a range is manually selected through the front panel or a remote command, this command is
automatically set to off.
Example
:SENS:CURR:RANG:AUTO ON
Set the range to be selected automatically for current
measurements.
Also see
(on page 7-67)