Section 2: General operation
Series 2280 Precision DC Power Supply Reference Manual
2-84
077085503 / March 2019
To save a user setup to external USB flash drive using SCPI commands:
Send the command:
:MMEMory:SAVE:SETup "mysetup"
Save the present setup to the external USB flash drive as
mysetup
.
Recall a user setup
You can recall setups from internal nonvolatile memory or external USB flash drive. When you recall
a setup, the instrument returns to the settings that are saved in memory.
To recall a saved setup from internal nonvolatile memory using the front panel:
1. Press the
MENU
key.
2. Under System, select
Save/Recall
.
3. Select the button next to
Recall
and press the
ENTER
key. The Save/Recall Location window is
displayed.
4. Select a saved setup in the pop-up window.
5. Press the
ENTER
key. The selected setup is applied to the instrument.
To recall a saved setup from external USB flash drive using the front panel:
1. Press the
MENU
key.
2. Under System, select
Save/Recall
.
3. Select the button next to
Recall
and press the
ENTER
key. The Save/Recall Location window is
displayed.
4. Select
Recall From USB
and press the
ENTER
key. The Save/Recall Location window is
displayed.
5. Select a saved setup in the pop-up window.
6. Press the
ENTER
key. The selected setup is applied to the instrument.
To recall a user setup from internal memory using SCPI commands:
*RCL <n>
Where
<n>
is an integer between 1 and 5 that represents the saved setup.
To recall a user setup from external USB flash drive using SCPI commands:
:MMEMory:LOAD:SETup "mysetup"
Recall the setup named
mysetup
in the external USB flash drive to the instrument.