Series 2280 Precision DC Power Supply Reference Manual
Section 2: General operation
077085503 / March 2019
2-83
Saving setups
You can save the present settings that you have defined for the Series 2280 to internal memory.
After the settings are saved, you can recall the settings. You can also set them to be the default
settings on power up.
Save a user setup
You can save a user setup to internal nonvolatile memory or an external USB flash drive.
To save a user setup to internal nonvolatile memory using the front panel:
1. Configure the Series 2280 to the settings that you want to save.
2. Press the
MENU
key.
3. Under System, select
Save/Recall
.
4. Select the button next to
Save
and press the
ENTER
key. The Save/Recall Location window is
displayed.
5. Select a location to save the setup and press
ENTER
. A keyboard is displayed.
6. Use the keyboard to enter the name of the setup.
7. Press the
ENTER
key. The setup is added to internal memory.
To save a user setup to external USB flash drive using the front panel:
1. Press the
MENU
key.
2. Under System, select
Save/Recall
.
3. Select the button next to
Save
and press the
ENTER
key. The Save/Recall Location window is
displayed.
4. Select
Save to USB
and press the
ENTER
key. A keyboard is displayed.
5. Use the keyboard to enter the name of the setup.
6. Press the
ENTER
key. The setup is saved to external USB flash drive.
To save a user setup to internal nonvolatile memory using SCPI commands:
Configure the instrument to the settings that you want to save. To save the setup, send the
command:
*SAV <n>
Where
<n>
is an integer between 1 and 5.
The setup saved through the
*SAV
command have the name
SETUP
x
, where
x
is the value you set
for
<n>
.