Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-40
077085503 / March 2019
Example
:CAL:PROT:STAT ON, 'KI002280'
Enable the calibration mode.
:CAL:PROT:STEP1
Start voltage DAC and ADC zero-calibration point.
:CAL:PROT:STEP1:DATA
0.2996
Enter the voltage reading from the external DMM.
*OPC?
Wait until step 1 finishes. When finished, this command
returns 1.
:CAL:PROT:STEP2
Start voltage DAC and ADC full-scale calibration.
:CAL:PROT:STEP2:DATA 29.945
Enter the voltage reading from external DMM.
*OPC?
Wait until step 2 finishes. When finished, this command
returns 1.
:CAL:PROT:STEP3
Start current ADC zero-point calibration on the 100 mA
range.
:CAL:PROT:STEP3:DATA
0.00505,5.05
Enter the voltage and resistance readings from the
external DMM.
*OPC?
Wait until step 3 finishes. When finished, this command
returns 1.
:CAL:PROT:STEP4
Start current ADC full-scale calibration on the 100 mA
range.
:CAL:PROT:STEP4:DATA
0.505,5.05
Enter the voltage and resistance readings from the
external DMM.
*OPC?
Wait until step 4 finishes. When finished, this command
returns 1.
:CAL:PROT:STEP5
Start current ADC zero-point calibration on the 10 A
range.
:CAL:PROT:STEP5:DATA
0.00298,0.0496
Enter the voltage and resistance readings from the
external DMM.
*OPC?
Wait until step 5 finishes. When finished, this command
returns 1.
:CAL:PROT:STEP6
Start current DAC zero-point calibration.
:CAL:PROT:STEP6:DATA
Wait for the instrument to collect data; you do not need to
enter any parameters.
*OPC?
Wait until step 6 finishes. When finished, this command
returns 1.
:CAL:PROT:STEP7
Start current ADC full-scale calibration on the 10 A range.
:CAL:PROT:STEP7:DATA
0.2996,0.0496
Enter the voltage and resistance readings from the
external DMM.
*OPC?
Wait until step 7 finishes. When finished, this command
returns 1.
:CAL:PROT:STEP8
Start current DAC full-scale calibration.
:CAL:PROT:STEP8:DATA
Wait for the instrument to collect data; you do not need to
enter any parameters.
*OPC?
Wait until step 8 finishes. When finished, this command
returns 1.
CAL:PROT:SAVE
Save calibration constants to nonvolatile memory.
CAL:PROT:STATe OFF,'KI002280'
Disable the calibration mode.