Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-124
077085503 / March 2019
:SYSTem:POSetup:STATe
This command enables or disables default setup loading. When it is set to
On
, the power setup loads setup 1 right
after it is powered on.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
0 (OFF)
Usage
:SYSTem:POSetup:STATe <state>
:SYSTem:POSetup:STATe?
<state>
Disable default setup loading:
OFF
or
0
Enable default setup loading:
ON
or
1
Example
:SYST:POS:STAT ON
:SYST:POS:STAT?
The default setup is configured to load
automatically when the instrument is turned on.
Also see
(on page 7-7)
:SYSTem:PRESet
This command returns the instrument to the factory default conditions and erases all user data stored in
nonvolatile memory.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:PRESet
Details
Sending this command will return the instrument to the factory default conditions and erases all user
data stored in nonvolatile memory, such as list parameter and setup information.
This command does not influence the status structure registers and error queue.
This command resets GPIB address and IP address to default settings. After sending this command,
make sure to use the new address to communicate with the instrument.
You should allow ample time for the command to be processed before attempting to communicate
with the instrument again.