Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-52
077085503 / March 2019
:INITiate[n][:IMMediate]
This command starts the trigger model.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:INITiate[n][:IMMediate]
[n]
Channel number: For single-channel instruments, the value is
1
Also see
(on page 3-25)
MMEMory subsystem
The commands in the MMEMory subsystem save or recall the setups to the external USB flash drive.
:MMEMory:LOAD:SETup
This command recalls the specified setup from the external USB flash drive to the instrument.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:MMEMory:LOAD:SETup <fileName>
<fileName>
A string that contains a name for the imported setup (maximum 8 characters)
Detail
This command allows you to recall the specified setup in the external USB flash drive to the
instrument.
Make sure that the external USB flash drive is connected to the instrument before you send this
command. An unstable connection between the USB flash drive and the instrument will cause an
error 520: "No USB flash drive found."