Series 2280 Precision DC Power Supply Reference Manual
Section 3: Functions and features
077085503 / March 2019
3-49
The following figure illustrates the process of measurement for sweep mode.
Figure 78: Sweep mode measurement time
Increasing the speed of a list
To increase the speed of a list:
Reduce the NPLC.
Turn autozero off. If autozero is on, the instrument takes new reference and zero values for every
reading. This can slow down sweep operation. Be aware that if you disable autozero,
measurements may drift and become erroneous. For more information on the autozero options,
(on page 4-5).
Aborting a list
A list can be stopped for the following reasons:
The limit set by the abort on limit setting was exceeded
The trigger model is aborted
You can stop the list while it is in progress. When you stop the list, all list commands in the trigger
model are terminated.