Series 2280 Precision DC Power Supply Reference Manual
Section 7: SCPI command reference
077085503 / March 2019
7-39
:CALibration:PROTected[:DC[n]]:STEP<step>
This command performs the calibration procedures.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:CALibration:PROTected[:DC[n]]:STEP1
:CALibration:PROTected[:DC[n]]:STEP2
:CALibration:PROTected[:DC[n]]:STEP3
:CALibration:PROTected[:DC[n]]:STEP4
:CALibration:PROTected[:DC[n]]:STEP5
:CALibration:PROTected[:DC[n]]:STEP6
:CALibration:PROTected[:DC[n]]:STEP7
:CALibration:PROTected[:DC[n]]:STEP8
[n]
Channel number: For single-channel instruments, the value is
1
<step>
The following steps represent the entire calibration process:
Step1: V-ADC and V-DAC zero-point calibration.
Step2: V-ADC and V-DAC full-scale calibration.
Step3: I-100 mA and I-10 mA ADC zero-point calibration.
Step4: I-100 mA and I-10 mA ADC full-scale calibration.
Step5: I-10 A and I-1 A ADC zero-point calibration.
Step6: I-DAC zero-point calibration.
Step7: I-10 A and I-1 A ADC full-scale calibration.
Step8: I-DAC full-scale calibration.
Details
This command is used to start each calibration step. Users should follow the calibration procedure.
See Example.