Series 2280 Precision DC Power Supply Reference Manual
Section 7: SCPI command reference
077085503 / March 2019
7-95
STATus subsystem
The STATus subsystem controls the status registers of the Series 2280. For additional information on
(on page D-1).
:STATus:MEASurement[:EVENt]?
This command reads and clears the Measurement Event Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:MEASurement[:EVENt]?
Details
This attribute reads the Measurement Event Register of the status model and then clears the register.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in
the register. For details, see
(on page D-17).
Also see
None
:STATus:MEASurement:ENABle
This command sets or reads the contents of the Measurement Event Enable Register of the status model.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:STATus:MEASurement:ENABle <n>
:STATus:MEASurement:ENABle?
[n]
The status of the Measurement Status Register
Details
This command sets or reads the contents of the enable register of the Measurement Event Register.
When one of these bits is set and the corresponding bit in the Measurement Event Register is set, the
MSB bit in the Status Byte Register is set.
When sending binary values, preface
<n>
with
#b
. When sending hexadecimal values, preface
<n>
with
#h
. No preface is needed when sending decimal values.