Section 3: Functions and features
Series 2280 Precision DC Power Supply Reference Manual
3-16
077085503 / March 2019
Using the SCPI commands:
:CALCulate2:FORMat <name>
Where <
name
> can be any of the following options separated by commas:
MINimum
MAXimum
MEAN
SDEViation
PKPK
For example, you could enable minimum and maximum calculations by sending:
:CALCulate2:FORMat MIN, MAX
Specifying the data in the buffer to perform with statistics
You must select either the voltage or current value in the buffer to perform statistics calculations.
When you configure these settings, it affects the Graph and Data Sheet screen and the outputs of the
following commands:
:CALCulate2:IMMediate?
:CALCulate2:DATA?
Using the front panel:
1. Press the
MENU
key.
2. Under Measure, select
Data Buffers
.
3. Select the button next to
Statistics Function
and press the
ENTER
key. The function selection
window is displayed.
4. Select
Voltage
or
Current.
5. Press the
Exit
key to leave the selection window.