HP3458A Emulation Mode
H-14
Description
The
OFORMAT
command controls the designated GPIB output format for readings transferred over
the bus for both normal readings and readings from memory.
The
OFORMAT
command is identical to the
MFORMAT
command. Both commands control the format
of both normal and memory readings.
Example
OFORMAT ASCII
Select ASCII format.
OPT?
Purpose
To request a report of installed options.
Format
OPT?
Response
0
No installed options
1
MEM1 or MEM2 extended memory option installed
Description
The OPT? query requests a report of installed options. The response is a numeric value that indicates
whether or not a memory option is installed.
PER
Purpose
To select the period measurement function.
Format
PER <max_input>, <%_resolution>
Parameter
<max_input>, <%_resolution>
Ignored
Description
The
PER
command selects the period function as the measurement mode for the Model 2002. Both
parameters are ignored as the Model 2002 uses only autoranging for period measurements.
The front panel does not display period measurements.
Example
PER
Select period function.
PRESET
Format
PRESET <type>
Parameter
<type>
NORM, FAST, or DIG
NORM
AZERO ON
NDIG 6
DCV AUTO
NPLC 1
DISP ON
NRDGS 1,AUTO
FSOURCE ACV
OCOMP OFF
TARM AUTO
TIMER 1
TRIG SYN