Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
Specifications are subject to change without notice.
Page 4 of 14
SPEC-2002 Rev. H / February 2009
Model 2002
Multimeter Specifications
AC VOLTS
ACV INPUT CHARACTERISTICS
Rms Range
Peak Input
Full Scale rms
Resolution
Input Impedance
Temperature Coefficient
23
±(% of r % of range) / °C
Outside T
CAL
±5°C
200 mV
1 V
210.0000
100 nV
1M
Ω
±2% with <140pF
0.004 + 0.001
2 V
8 V
2.100000
1 µV
1M
Ω
±2% with <140pF
0.004 + 0.001
20
V 100
V 21.00000 10
µV 1M
Ω
±2% with <140pF
0.006 + 0.001
200 V
800 V
210.0000
100 µV
1M
Ω
±2% with <140pF
0.006 + 0.001
750 V
1100 V
775.000
1 mV
1M
Ω
±2% with <140pF
0.012 + 0.001
AC Voltage Uncertainty
= ±[ (% of reading) x (measured value) + (% of range ) x (range used) ] / 100.
PPM Accuracy
= (% accuracy) x 10,000.
0.015% of Range
= 30 counts for ranges up to 200V and 113 counts on 750V range at 5H digits.
LOW FREQUENCY MODE RMS
24
90 Days, 1 Year or 2 Years, ±2°C from last AC self-cal, for 1% to 100% of range
25
, ±(% of r %
of range)
Range 1–10Hz
22
10–50Hz 50–499Hz 501Hz–2kHz 2–10kHz
10–30kHz 30–50kHz 50–100kHz 100–200kHz 0.2–1MHz 1–2MHz
200 mV
0.09+0.015 0.06+0.015
0.035+0.015
0.03+0.02
0.02+0.02
0.025+0.02 0.05+0.02 0.3+0.015 0.75+0.025 2+0.1 5+0.2
2
V
0.09+0.015
0.04+0.015
0.025+0.015 0.02+0.02 0.02+0.02 0.025+0.02 0.05+0.02 0.3+0.015 0.75+0.025 2+0.1 5+0.2
20 V
0.1 +0.015 0.06+0.015
0.035+0.015
0.03+0.015
0.04+0.015
0.05 +0.015
0.07+0.015
0.3+0.015
0.75+0.025
4+0.2
7+0.2
22
200 V
21
0.1 +0.015 0.05+0.015
0.03 +0.015
0.03+0.015
0.04+0.015
0.05 +0.015
0.07+0.015
0.3+0.015
0.75+0.025
22
4+0.2
22
750 V
21
0.13+0.015 0.09+0.015
0.05 +0.015
0.05+0.015
0.06+0.015
0.08 +0.015
0.1 +0.015
22
0.5+0.015
22
NORMAL MODE RMS
24
90 Days, 1 Year or 2 Years, ±2°C from last AC self-cal, for 1% to 100% of range
25
, ±(% of r % of range)
Range 20–50Hz 50–100Hz
0.1–2kHz 2–10kHz 10–30kHz 30–50kHz
50–100kHz
100–200kHz 0.2–1MHz 1–2MHz
200 mV
0.25+0.015
0.07+0.015
0.02+0.02
0.02+0.02
0.025+0.02
0.05+0.02
0.3+0.015
0.75+0.025
2+0.1
5+0.2
2 V
0.25+0.015
0.07+0.015
0.02+0.02
0.02+0.02
0.025+0.02
0.05+0.02
0.3+0.015
0.75+0.025
2+0.1
5+0.2
20
V
0.25+0.015 0.07+0.015 0.03+0.015 0.04+0.015 0.05
+0.015 0.07+0.015 0.3+0.015
0.75+0.025
4+0.2
7+0.2
22
200 V
21
0.25+0.015 0.07+0.015 0.03+0.015 0.04+0.015 0.05
+0.015 0.07+0.015 0.3+0.015 0.75+0.025
22
4+0.2
22
750 V
21
0.25+0.015
0.1 +0.015
0.05+0.015
0.06+0.015
0.08 +0.015
0.1 +0.015
22
0.5+0.015
22
dB ACCURACY RMS
±dB, 90 Days, 1 Year or 2 Years, TCAL ±5°C, Reference=1V, Autoranging, Low Frequency Mode, AC+DC Coupling
Input 1–100Hz
0.1–30kHz
30–100kHz 100–200kHz 0.2–1MHz 1–2MHz
–54 to –40 dB (2 mV to 10 mV)
0.230
0.225
0.236
0.355
–40 to –34 dB (10 mV to 20 mV)
0.036
0.031
0.041
0.088
–34 to 6 dB (20 mV to 2 V)
0.023
0.018
0.028
0.066
0.265
0.630
6 to 26 dB (2 V to 20 V)
0.024
0.024
0.028
0.066
0.538
0.820
22
26 to 46 dB (20 V to 200 V)
0.024
0.024
0.028
0.066
22
0.538
22
46 to 57.8 dB (200 V to 775 V)
0.018
0.021
0.049
22
ACV READING RATES
17, 22
Readings/Second to Memory
Readings/Second to IEEE-488
18
Readings/Second with
Time Stamp to IEEE-488
18
PLC
Measurement
Aperture
Bits
Default
Digits
Autozero Off
Autozero On
Autozero Off
Autozero On
Autozero Off
Autozero On
10
167 ms (200 ms)
29
6H
6 (5)
2 (1.7)
6 (5)
2 (1.6)
6 (5)
2 (1.6)
2
33.4 ms (40 ms)
27
5H
29 (25)
9 (7.6)
28 (23)
9 (7.4)
26 (21)
9 (7.4)
1
16.7 ms (20 ms)
26
5H
56 (48)
47 (40)
52 (43)
44 (36)
48 (39)
40 (33)
0.2
3.34 ms (4 ms)
23
5H
145 (129)
110 (98)
131 (117)
100 (88)
102 (91)
79 (70)
0.1
1.67 ms (2 ms)
22
5H
150 (144)
112 (108)
132 (127)
101 (97)
102 (98)
80 (77)
0.02
334 µs (400 µs)
20
5H
150 (150)
115 (115)
132 (132)
103 (103)
102 (102)
80 (80)
0.01
167 µs (167 µs)
19
4H
382 (382)
116 (116)
251 (251)
103 (103)
163 (163)
80 (80)
0.01
13
167 µs (167 µs)
19
4H
2000 (2000)
2000 (2000)
23
Temperature coefficient applies to rms and average readings. For frequencies above 100kHz, add 0.01% of reading/°C to temperature coefficient.
24
Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, autozero on, digital filter off, following 55-minute warm-up.
25
For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% of range uncertainty. For inputs from 200kHz to 2MHz, specifications