IEEE-488 Reference
3-99
:REFerence <n>
[:SENSe[1]]:CURRent:AC:REFerence <n>
Specify reference for ACI
[:SENSe[1]]:CURRent:DC:REFerence <n>
Specify reference for DCI
[:SENSe[1]]:VOLTage:AC:REFerence <n>
Specify reference for ACV
[:SENSe[1]]:VOLTage:DC:REFerence <n>
Specify reference for DCV
[:SENSe[1]]:RESistance:REFerence <n>
Specify reference for
Ω
2
[:SENSe[1]]:FRESistance:REFerence <n>
Specify reference for
Ω
4
[:SENSe[1]]:FREQuency:REFerence <n>
Specify reference for FREQ
[:SENSe[1]]:TEMPerature:REFerence <n>
Specify reference for TEMP
<n> = -2.1 to +2.1
Reference for ACI and DCI
-1100 to +1100
Reference for DCV and ACV
0 to 1.05e9
Reference for
Ω
2
0 to 2.1e6
Reference for
Ω
4
0 to 1.5e7
Reference for FREQ
-328 to +3310
Reference for TEMP (°F)
-200 to +1821
Reference for TEMP (°C)
73 to 2094
Reference for TEMP (K)
DEFault
0 (All functions)
MINimum
Minimum value for specified function
MAXimum
Maximum value for specified function
:REFerence?
Query programmed reference value
:REFerence? DEFault
Query *RST default reference value
:REFerence? MINimum
Query lowest allowable reference value
:REFerence? MAXimum
Query largest allowable reference value
These commands are used to establish a reference value for the specified function. When Ref-
erence is enabled (see :REFerence:STATe), the result will be the algebraic difference between
the input signal and the reference value:
Reading = Input signal - Reference
From the front panel, reference is called relative (REL). See paragraph 2.6 for more information
on REL.
The :REFerence <n> command is coupled to the :ACQuire command. The last command sent
(:REFerence <n> or :ACQuire) establishes the reference. When a reference is set using the
:REFerence <n> command, the :REFerence? query command returns the programmed value.
Conversely, when a reference is set using the :ACQuire command, the :REFerence? query com-
mand returns the acquired reference value.
PRINT #1, "output 16; :curr:ac:ref 1; ref?"
' Set reference to 1A
PRINT #1, "enter 16"
' Get response from 2002
3.18.9
Parameters
Query
Description
Program fragment
Summary of Contents for 2002
Page 123: ...Front Panel Operation 2 106...
Page 145: ...IEEE 488 Reference 3 22...
Page 167: ...IEEE 488 Reference 3 44...
Page 284: ...A Specifications A 1...
Page 362: ......
Page 364: ...This page left blank intentionally...
Page 365: ......