IEEE-488 Reference
3-148
:POINts <n>
:TRACe:POINts <n>
Specify buffer size
No memory
option
MEM 1
MEM 2
<n> =
2 to 404
2 to 1381
2 to 5980
(Full)
2 to 2027
2 to 6909
2 to 29908
(Compact)
MAXimum
404
1381
5980
(Full)
2027
6909
29908
(Compact)
DEFault
100 (all cases)
MINimum
2 (all cases)
:POINts?
This command is used to specify the size of the buffer. Notice that the valid parameter range
depends on which memory option is installed, and which element group (full or compact) is
being used (see :EGRoup).
PRINT #1, "output 16; :trac:poin 50; poin?"
' Set size of buffer
PRINT #1, "enter 16"
' Get response from 2002
:AUTO <b>
:TRACe:POINts:AUTO <b>
Control auto buffer sizing
<b> = 0 or OFF
Disable auto buffer sizing
1 or ON
Enable auto buffer sizing
:AUTO?
This command is used to enable or disable auto buffer sizing. When enabled, the buffer size
defaults to the measure count value in the Trigger Model (see :TRIGger:COUNt in Trigger sub-
system). With auto buffer enabled, changing the Trigger Model’s measure count value changes
the buffer size to that value.
Disabling auto buffer sizing does not affect the size of the buffer. It simply means that you must
then use the :POINts command to change the buffer size.
If you use the :POINts command to change the buffer size while auto is enabled, the buffer size
changes to that value and auto buffer sizing disables.
If you try to enable auto buffer sizing while the measure count is programmed for infinite, an
error will occur and the command will be ignored. You cannot have an infinite buffer size.
PRINT #1, "output 16; :trig:coun 200"
' Set measure count
PRINT #1, "output 16; :trac:poin:auto on; auto?"
' Enable auto buffer sizing
PRINT #1, "enter 16"
' Get response from 2002
:FEED <name>
:TRACe:FEED <name>
Specify readings source
<name> = SENSe[1]
Put raw readings in buffer
CALCulate[1]
Put calculated readings in buffer
NONE
Put no readings in buffer
3.22.4
Parameters
Query
Description
Program fragment
Parameters
Query
Description
Program fragment
3.22.5
Parameters
Summary of Contents for 2002
Page 123: ...Front Panel Operation 2 106...
Page 145: ...IEEE 488 Reference 3 22...
Page 167: ...IEEE 488 Reference 3 44...
Page 284: ...A Specifications A 1...
Page 362: ......
Page 364: ...This page left blank intentionally...
Page 365: ......