IEEE-488 Reference
3-102
:AUTO <b>|ONCE
[:SENSe[1]]:CURRent:AC:DIGits:AUTO <b>|ONCE
Control auto resolution for ACI
[:SENSe[1]]:CURRent:DC:DIGits:AUTO <b>|ONCE
Control auto resolution for DCI
[:SENSe[1]]:VOLTage:AC:DIGits:AUTO <b>|ONCE
Control auto resolution for ACV
[:SENSe[1]]:VOLTage:DC:DIGits:AUTO <b>|ONCE
Control auto resolution for DCV
[:SENSe[1]]:RESistance:DIGits:AUTO <b>|ONCE
Control auto resolution for
Ω
2
[:SENSe[1]]:FRESistance:DIGits:AUTO <b>|ONCE
Control auto resolution for
Ω
4
[:SENSe[1]]:TEMPerature:DIGits:AUTO <b>|ONCE
Control auto resolution for TEMP
<b> =
1 or ON
Enable auto resolution
0 or OFF
Disable auto resolution
ONCE
Enable and then disable auto resolution
:AUTO?
Query state of auto resolution
These commands are used to enable or disable auto resolution for the specified function. With
auto resolution selected, the instrument automatically selects the optimum resolution for the
present integration time setting (see Tables 2-10, 2-15, 2-20 and 2-25).
The ONCE parameter is analogous to a momentary toggle switch. When :AUTO ONCE is sent,
auto resolution turns on momentarily and then disables. While enabled, resolution is selected
automatically. Auto resolution then disables but the selected value is retained.
The auto resolution command (:DIGits:AUTO) is coupled to the command that manually selects
resolution (:DIGits <n>). When auto resolution is enabled, the parameter value for :DIGits
<n> changes to the automatically selected resolution value. Thus, when auto resolution is dis-
abled, the instrument remains at the automatically selected resolution.
PRINT #1, "output 16; :volt:dc:dig:auto on; auto?"
' Enable auto resolution
PRINT #1, "enter 16"
' Get response from 2002
:AVERage commands
The :AVERage commands are used to configure and control the filter. See paragraph 2.9 for
details.
:TCONtrol <name>
[:SENSe[1]]:CURRent:AC:AVERage:TCONtrol <name>
Select filter type for ACI
[:SENSe[1]]:CURRent:DC:AVERage:TCONtrol <name> Select filter type for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:TCONtrol <name>
Select filter type for ACV
[:SENSe[1]]:VOLTage:DC:AVERage:TCONtrol <name>
Select filter type for DCV
[:SENSe[1]]:RESistance:AVERage:TCONtrol <name>
Select filter type for
Ω
2
[:SENSe[1]]:FRESistance:AVERage:TCONtrol <name>
Select filter type for
Ω
4
[:SENSe[1]]:TEMPerature:AVERage:TCONtrol <name>
Select filter type for TEMP
<name> = REPeat
Select repeating filter
MOVing
Select moving filter
:TCONtrol?
Query filter type
These commands are used to select the type of averaging filter (REPeat or MOVing) for the
specified function.
Parameters
Query
Description
Program fragment
3.18.11
Parameters
Query
Description
Summary of Contents for 2002
Page 123: ...Front Panel Operation 2 106...
Page 145: ...IEEE 488 Reference 3 22...
Page 167: ...IEEE 488 Reference 3 44...
Page 284: ...A Specifications A 1...
Page 362: ......
Page 364: ...This page left blank intentionally...
Page 365: ......