Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
Specifications are subject to change without notice.
Page 14 of 14
SPEC-2002 Rev. H / February 2009
Model 2002
Multimeter Specifications
Physical
Case Dimensions:
90mm high
×
214mm wide
×
369mm deep (3½ in.
×
8½ in.
×
14½ in.).
Working Dimensions:
From front of case to rear including power cord and IEEE-488 connector: 15.0 inches.
Net Weight:
<4.2kg (<9.2 lbs.).
Shipping Weight:
<9.1kg (<20 lbs.).
Standards
EMI/RFI:
Conforms to VDE 0871B (per Vfg 1046/1984), IEC 801-2. Meets FCC part 15 Class B, CISPR-22 (EN55022).
Safety:
Conforms to IEC348, CAN/CSA-C22.2. No. 231, MIL-T-28800E
1
. Designed to UL1244.
Accessories Supplied
The unit is shipped with line cord, high performance modular test leads, operator’s manual, option slot cover, and
full calibration data.
EXTENDED MEMORY/NON-VOLATILE MEMORY OPTIONS
DATA STORAGE
Setup Storage
Model Size
(Bytes)
4½-Digit
6½-Digit with
Time Stamp
Type
Number Type
2002 8k
2,027 404
volatile
1 non-volatile
2002/MEM1 32k 6,909
1,381
non-volatile
5
non-volatile
2002/MEM2 128k 29,908
5,980
non-volatile
10
non-volatile
These are the minimum sizes to expect.