HP3458A Emulation Mode
H-15
FAST
Same as NORM with the following exceptions:
DCV 10
OFORMAT DINT
DISP OFF
TRIG AUTO
AZERO OFF
DIG:
Not supported.
Default
NORM
Description
The
PRESET
command selects one of two preset instrument configurations, one for normal opera-
tion, and one for fast operation. The DIG preset configuration supported by the HP3458A is not sup-
ported by the Model 2002.
Example
PRESET FAST
Select fast configuration.
RANGE
Purpose
To select the measurement range.
Format
RANGE <max_input> ,<%_resolution>
Parameter
<max_input>
Manual range or AUTO (autorange)
<%_resolution>
Ignored
Default
AUTO
Query
RANGE?
Returns the present measurement range for the current function.
Description
The
RANGE
command allows you to select a manual measurement range or autorange. (See paragraph
3.18 for range information specific to each function.) Specifying AUTO for the
<max_input>
parameter enables autoranging, and the
<%_resolution>
parameter is ignored.
Example
RANGE AUTO
Select autorange.
RESET
Purpose
To reset the Model 2002 to the power-on state.
Format
RESET
Parameter
None
Description
The
RESET
command resets the multimeter to the power-on state without cycling power. Reset
effectively executes: