HP3458A Emulation Mode
H-18
<type> parameter
Numeric equivalent
Description
ANA
1
RMS conversion
RNDM
2
Low-frequency voltage measurements
SYNC
3
Low-frequency voltage measurements
Default
ANA
Query
SETACV?
Returns the currently specified conversion technique. A “3” is returned for RNDM
or SYNC.
Description
The
SETACV
command selects the RMS conversion technique to be used for the AC or AC+DC volt-
age measurement modes. Note that both RNDM and SYNC choose the low-frequency voltage mea-
surement mode.
Example
SETACV SYNC
Select low-frequency AC mode.
SRQ
Purpose
To enable SRQ.
Format
SRQ
Parameter
None
Description
The
SRQ
command sets bit 2 in the Model 2002 status register. When bit 2 is enabled, the SRQ line
will be asserted when certain conditions occur. (See
RQS
command.)
SSTATE
Purpose
To store the instrument's present configuration in memory.
Format
SSTATE <state>
Parameter
<state>
Memory option
Valid <state> parameters
No extended memory
SAV0
2002-MEM1 installed
SAV0-SAV4
2002-MEM2 installed
SAV0-SAV9
Description
SSTATE
stores the multimeter’s present state in the given state memory location. Note that the num-
ber of available states depends on installed memory options as indicated above. Instrument states
may recalled with the
RSTATE
command.
Example
SSTATE SAV0
Save state 0.