Model 2002
Multimeter Specifications
ACV CREST FACTOR MEASUREMENT
26
Crest Factor:
= Peak AC / rms AC.
Crest Factor Resolution:
3 digits.
Crest Factor Accuracy:
Peak AC uncer AC normal mode rms uncertainty.
Measurement Time:
100ms plus rms measurement time.
Input Characteristics:
Same as ACV input.
Crest Factor Frequency Range:
20Hz – 1MHz.
Crest Factor Display:
Access as multiple display on AC volts.
AC COUPLING
For AC only coupling, add the following % of reading:
1–10Hz 10–20Hz 20–50Hz 50–100Hz
100–200Hz
Normal Mode (rms, average)
—
—
0.41
0.07
0.015
Low Frequency Mode (rms)
0.1
0.01
0 0 0
For low frequency mode below 200Hz, specifications apply for sine wave inputs only.
AC+DC COUPLING
For DC >20% of AC rms voltage, apply the following additional uncertainty, multiplied by the ratio (DC/total rms). Applies to rms and average
measurements.
Range
% of Reading
% of Range
200mV, 20V
0.05
0.1
2V, 200V, 750V
0.07
0.01
AVERAGE ACV MEASUREMENT
Normal mode rms specifications apply from 10% to 100% of range, for 20Hz–1MHz. Add 0.025% of range uncertainty for 50kHz–100kHz, 0.05%
of range uncertainty for 100kHz–200kHz, and 0.5% of range uncertainty for 200kHz–1MHz.
HIGH CREST FACTOR ADDITIONAL ERROR ±(% OF READING)
Applies to rms measurements.
Crest Factor
1 – 2
2 – 3
3 – 4
4 – 5
Additional Error
0 0.1 0.2 0.4
ACV PEAK VALUE MEASUREMENT
19
Repetitive Peak Accuracy, ±(% of % of range), 90 Days, 1 Year or 2 Years, T
CAL
±5°C
Range
20Hz–
1kHz
27
1kHz–
10kHz
10kHz–
30kHz
30kHz–
50kHz
50kHz–
100kHz
100kHz–
300kHz
300kHz–
500kHz
500kHz–
750kHz
750kHz–
1MHz
Temperature Coefficient
±(% of r% of range)/°C
Outside T
CAL
±5°C
200 mV
0.08+0.7
0.09+0.7
0.1 +0.7
0.15+0.7
0.25+0.7
1.0+0.7
2.5+0.7
5.5+0.7
9+0.7
0.002 + 0.03
2 V
0.08+0.3
0.09+0.3
0.1 +0.3
0.15+0.3
0.25+0.3
1.0+0.3
2.5+0.3
5.5+0.3
9+0.3
0.002 + 0.03
20 V
0.1 +0.7
0.11+0.7
0.14+0.7
0.19+0.7
0.25+0.7
1.0+0.7
2.5+0.7
5.5+0.7
9+0.7
0.004 + 0.03
200 V
21
0.1 +0.3
0.11+0.3
0.14+0.3
0.19+0.3
0.25+0.3
1.0+0.3
22
2.5+0.3
22
5.5+0.3
22
9+0.3
22
0.004 + 0.03
750 V
21
0.12+0.6
0.16+0.6
0.2 +0.6
0.25+0.6
22
0.5
+0.6
22
0.01
+
0.02
Valid % of Range
28
10–400% 10–400% 10–400% 10–350% 10–350% 10–250% 10–150% 10–100% 7.5–75%
Default Measurement Resolution:
4 digits.
Non-Repetitive Peak:
10% of range per µs typical slew rate for single spikes.
Peak Width:
Specifications apply for all peaks
≥
1µs.
Peak Measurement Window:
100ms per reading.
Maximum Input:
±1100V peak, 2 x 1028V·Hz (for inputs above 20V).
26
Subject to peak input voltage specification.
27
AC peak specifications assume AC + DC coupling for fre-quencies below 200Hz.
28
For overrange readings 200–300% of range, add 0.1% of reading uncertainty. For 300–400% of range, add 0.2% of reading uncertainty.