Selection of the Measuring Point
Inlet Outlet
L
≥
10 D
L
≥
5 D
Disturbance source
: 2 x 90 °-elbows in one plane
Inlet Outlet
L
≥
25 D
L
≥
5 D
Disturbance source
: 2 x 90 °-elbows in different planes
Inlet Outlet
L
≥
40 D
L
≥
5 D
Disturbance source
: T-section
Inlet Outlet
L
≥
50 D
L
≥
10 D
Disturbance source
: diffuser
Inlet Outlet
L
≥
30 D
L
≥
5 D