Process Interfaces
13 - 2
ADM 6725 E10
KA
T
Hold last value
Error-value
!
S9-5f2
I [mA]
20
4
t
Maximum(20.00mA)
Error-value
!
S9-5f3
I [mA]
20
4
t
t
d
Other value...
Error-value
!
2.00 mA
Error-value:
S9-5f4
I [mA]
20
4
t
t
d
Error-Value Delay
Normally, FLUXUS uses the value of the entered damping factor for the error-value delay
with respect to the process outputs. However, you can also input a specific value for this
delay by the following method. You can enable/disable the error-value delay prompt in
program branch SPECIAL FUNCTIONS
(see chapter 7.7.2)
. Additionally, it can be
enabled/disable it with the HotCode method
(see chapter 2.4.4)
:
NEXT
M UX
Enter this key sequence from the MAIN
MENU after pressing key BRK.
>DAMPING< edit
Error-val.delay
S9-2e The current status is displayed:
>DAMPING< indicates that FLUXUS uses
the damping factor as the time for the delay.
If you select >EDIT<, a specific error-delay
value can be set.
s
Error-val.delay
O27
Enter the time delay required.
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Page 4: ...ADM 6725 E10 KAT ...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...