Process Interfaces
ADM 6725 E10
KA
T
13 - 3
13.1.2 Set-up of Current Loop
Note:
The display 'Current Loop'
only
appears in program branch OPT
if
FLUXUS is
equipped with a current loop, which has been installed in program branch SPECIAL
FUNCTIONS (see chapter 7.7.5).
no >YES<
Current Loop
O12
Select
YES
, if you wish to activate the current
loop analogue output.
13.1.3 Set-up of Voltage Output
Note:
The display ‘Voltage Output’
only
appears in program branch OPT
if
FLUXUS is
equipped with a voltage output, which has been installed in program branch
SPECIAL FUNCTIONS (see chapter 7.7.5).
no >YES<
Voltage Output
O25
Select
YES
, if you wish to activate the
analogue voltage output.
13.1.4 Set-up of Frequency Output
Note:
The display 'Frequency Output'
only
appears in program branch OPT
if
FLUXUS is
equipped with a frequency output, which has been installed in program branch
SPECIAL FUNCTIONS (see chapter 7.7.5).
no >YES<
Frequency Output
O14
Select
YES
, if you wish to activate the
analogue frequency output.
13.1.5 Set-up of the Full-Scale Value for all Analogue
Outputs
FLUXUS will display the unit of measurement
(m3/h, in our example)
according to the
previously selected quantity of measurement (in this case: volume flow).
0.1 m3/h
Full-Scale Val.
O13
b
O15
b
O26
b
This display requires the input for the full-scale
value of the respective analogue output. Specify
as 'Full-Scale Val.' the highest measuring value,
which can be expected. The entered parameter
for the full-scale value would correspond to 100 %
of the measured value = 20 mA, 1 V or 10 V,
1000 Hz.
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Page 4: ...ADM 6725 E10 KAT ...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...