ADM 6725 E10
KA
T
14 - 3
Errors
NO COUNTING !
Reason:
You activated the total quantity counters, although
flow velocity was selected as the quantity of
measurement. This is not possible.
Corrective action:
Acknowledge message by pressing ENTER. Should
you really want to know the value of total quantity
throughput, select a quantity of measurement
different from the flow velocity.
14.1.3 Errors during Data Transfer of Measured Values
Error message F9:
NO DATA !
Print Meas.Val.
Reason:
Special function 'Print Meas. Val.' was activated, al-
though no measured values are stored in FLUXUS.
Maybe, you did not select 'Store Meas. Data' in the
program branch OUTPUT OPTIONS.
(See chapter 6.5
'Storage of Measured Values with Internal Data Logger'.)
Corrective action:
Activate 'Store Meas. Data' and repeat measurement.
Error message F10:
SERIAL ERROR !
Print Meas.Val.
Reason:
There is a problem with the serial communication. It
might be e.g., that the receiving device is not ready or
not yet connected.
Corrective action:
Check connections and check whether the receiving
device is ready to receive data.
Error message F11:
PRINTER TIMEOUT!
>RETRY< Abort
Reason:
One of the special functions 'List Meas. Sets' or
'Graphic Meas. Sets' was selected, although the print-
er is not ready yet.
Corrective action:
Check printer.
Summary of Contents for FLUXUS ADM 6725
Page 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Page 4: ...ADM 6725 E10 KAT ...
Page 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Page 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Page 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Page 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Page 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...