7.1 Warning and error messages UL3000/UL6000
Type
Notification
Possible sources of
error
Remedy
W102
Timeout during
communication with
EEPROM on VI-Board
The EEPROM on the
VI-Board is defective or
does not exist
• Contact customer service
W104
One EEPROM parameter
has been initialized
A new parameter was
introduced by a
software update
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameter corresponds to your
application
The EEPROM on the
VI-Board is defective
• Confirm the warning message
• Check if the message occurs each time
when you restart the device
• Contact customer service
W106
Several EEPROM
parameters have been
initialized
A software update
introduced new
parameters
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameters corresponds to your
application
The EEPROM on the
VI-Board was replaced
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameters corresponds to your
application
The EEPROM on the
VI-Board is defective
• Confirm the warning message
• Check if the message occurs each time
when you restart the device
• Contact customer service
E107
Internal IIC communication
error
Internal IIC
communication error
• Contact customer service
E108
Internal IIC2 communication
error
Internal IIC2
communication error
• Contact customer service