7 | Warning and error messages
INFICON
100 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
Type
Notification
Possible sources of
error
Remedy
W208
24V fan supply voltage out
of range
Malfunction of a fan
• Contact customer service
Short circuit or
overload in the 24V fan
supply
• Contact customer service
W209
24V DIV1 supply voltage out
of range
Short circuit or
overload in the 24V
DIV1 supply
• Contact customer service
W210
24V DIV2 supply voltage out
of range
Short circuit or
overload in the 24V
DIV2 supply
• Contact customer service
W211
5V internal supply voltage
out of range
Short circuit or
overload in the internal
5V supply
• Contact customer service
W212
5V sniffer line supply voltage
out of range
Sniffer line defective
• Use another sniffer line, if possible
• Contact customer service
Short circuit or
overload in the 5V
sniffer line supply
• Contact customer service
W221
Internal voltage 24V_RC
voltage out of range
The remote control is
defective
• Use another remote control, if possible
The cable connected to
REMOTE CONTROL
connector is defective
• Use another cable, if possible
The accessory
connected to
ACCESSORIES
connector is defective
• Use another accessory, if possible
The cable connected to
ACCESSORIES
connector is defective
• Use another cable, if possible
Short circuit or
overload in the
24V_RC supply
• Contact customer service