Digital input pin 5 (IO1000)
ZERO
Digital input pin 6 (IO1000)
CAL
Digital input pin 7 (IO1000)
Clear
Digital input pin 8 (IO1000)
Purge
Digital input pin 9 (IO1000)
Start/Stop
Digital input pin 10 (IO1000)
No function
Machine factor vacuum mass 2
1
Machine factor vacuum mass 3
1
Machine factor vacuum mass 4
1
Leak rate filter
I•CAL
Cathode
Cathode 1 auto (can only be changed
by service)
Lower pressure limit SL200 capillary
clogged
0.1 mbar
Upper pressure limit SL200 capillary broken
1 mbar
Audio alarm
TRIGGER
Audio alarm delay
60 s
Notification beep
On
Inlet area offset mass 2
1,3×10
-10
A (Modifiable by function
"Determine the background of the inlet
74]")
Inlet area offset mass 3
6,7×10
-14
A (Modifiable by function
"Determine the background of the inlet
74]")
Inlet area offset mass 4
3,3×10
-15
A (Modifiable by function
"Determine the background of the inlet
74]")
Analog output IO1000 upper exponent
-5
Pressure unit (display)
mbar
Leak rate unit vacuum operation (display)
mbar l/s
Leak rate unit sniffer operation (display)
mbar l/s
Sniffer line light alarm configuration (SL3000
only)
Flashing
Sniffer line light brightness (SL3000 only)
5
Sniffer line beep (SL3000 only)
Beep when setpoint 1
Operation mode
Vacuum
Amplifier test during calibration
On (can only be changed by service)
Alternating preamplifier ranges monitoring
On (can only be changed by service)