Warning and error messages | 7
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
103 / 152
Type
Notification
Possible sources of
error
Remedy
W311
Cathode 2 broken
Cathode defective
• Check that the other cathode has been
switched on
• Contact customer service
Connection to cathode
broken
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
E312
Cathodes broken
Both cathodes
defective
• Contact customer service
Connection to cathode
broken
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
E340
Emission error
Pressure too high with
aged cathode
• Contact customer service
W342
Cathodes not connected
Both cathodes
defective during self-
test
• Contact customer service
Plug of ion source not
connected
• Contact customer service
W350
Suppressor not connected
Suppressor cable not
connected or defective
• Contact customer service
E352
Preamplifier not connected
Preamplifier defective
or cable not connected
• Contact customer service
W358
Alternate preamplifier
ranges
Preamplifier defective
• Contact customer service
Unusual leak rate
waveform
• Contact customer service
W359
Overdriven preamplifier
Mass spectrometer
contaminated
• Check the tightness of the connection
between the leak detector and the test
object
• Reduce contamination in the environment
where the device is located
• Use the 'Regeneration' function in the
'Vacuum' operation mode
Ion source or
preamplifier defective
• Contact customer service