8 | Cleaning and maintenance
INFICON
120 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
WARNING
Risk of injury due to improper mounting of a gas cylinder on the mobile
leak detector
► When mounting the bottle holder on the leak detector, follow the sequence of action
steps as described in the separate mounting instructions.
► Mount the base of the cylinder holder flush with the gas cylinder.
► In order not to impair the stability of the leak detector, transport only small to
medium-sized cylinders (maximum 10 l, maximum cylinder size including protective
cap 98 cm, 200 bar).
► Check the supplied tensioning straps for visible damage.
► Always secure the bottle on the bottle holder with both tension straps.
► Make sure that any connected spray hose is never under tension.
► Observe the manufacturer's safety instructions for handling gas cylinders.
8.7 Creating screenshots
You can save the current screen contents of the device to an image file. You can use
such a file for service communication, for example.
1
Create a directory named "Screenshots" on a USB stick (FAT 32 formatted).
2
To take a screenshot, connect the USB stick to one of the USB ports of the leak
detector, see also the back of the “Control unit [
ð
A screenshot is automatically taken and stored in the directory on the USB
stick. The date and time are also saved.
3
To take another screenshot, disconnect the USB stick from the USB port and
then reconnect it to the leak detector.
ð
A screenshot saved earlier will not be overwritten when another screenshot is
taken.
8.8 Have the device serviced or repaired
WARNING
Danger due to harmful substances
Contaminated devices could endanger health. The contamination declaration serves
to protect all persons who come into contact with the device.
► Fill in the declaration of contamination completely.
Maintenance inside the device should only be performed by the manufacturer. We
recommend having the device serviced every 4000 hours or annually.
Summary of Contents for UL6000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 125 152 ...
Page 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 ...
Page 151: ......
Page 152: ......