6 | Operation
INFICON
84 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
Returns measured values and metadata of the measurement cycle with ID 4 in the
format “csv_en”.
Example 2:
http://ul3xxx.inficon.com:3000/md?dts=2018-11-05T23:00:00.000Z&dte=
2018-11-06T23:00:00.000Z&f=json&d=r
Returns measured values between 2018-11-05T23:00:00.000Z und
2018-11-06T23:00:00.000Z in the “json” format.
Parameter Name
Description
Options
Example
c
Measureme
nt cycle
Querying one or more
measurement cycles via
ID
ID's of the
measurement cycles.
Separated by comma
for several
c=3
c=3, 5, 8
dts
START
The start time of the time
period from which data is
to be exported.
A date in ISO format
in UTC time
dts=2018-08-02T17:15:12.000Z
dte
End
The end time of the time
period from which data is
to be exported.
A date in ISO format
in UTC time
dte=2018-08-02T17:45:35.000Z
f
Format
The data format of the
export
json, csv_en, csv_de
Standard: csv_en
f=json
f=csv_en
d
Scope
Scope of the data. With
or without metadata
mr: Meta and
measured values
m: Meta only
r: only measured
values
Standard: mr
d=mr
d=r
6.7.4 Measurement data database: Information
The maximum size of the database corresponds approximately to the number of
measured values that are generated over the period of one month during continuous
recording. This corresponds to a database size of approx. 1.5 GB.
If storage is needed for the recording and the maximum database size (in
measurement values) has been exceeded, the oldest records are automatically
deleted so that 5% of the specified capacity is freed for new data.
ü
Supervisor
rights
►
> Measurement data > Database information
ð
The following is displayed:
"Current database size" in [Byte]
“Measurement values“: Number of measurement values in the database
“Max. database size": Number of measurement values that can be stored in
Summary of Contents for UL6000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 125 152 ...
Page 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 ...
Page 151: ......
Page 152: ......