7 | Warning and error messages
INFICON
102 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
Type
Notification
Possible sources of
error
Remedy
W254
Valve current is out of range
One or more lowered
valves are not
connected or defective
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
W255
Fan current is out of range
Fan not connected
• Contact customer service
Fan blocked or
defective
• Contact customer service
W300
Anode voltage too low
Short circuit of the
anode voltage
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
W301
Anode voltage too high
MSB defective
• Contact customer service
W302
Suppressor voltage too low
Short circuit of
suppressor or
preamplifier
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
W303
Suppressor voltage too high
MSB defective
• Contact customer service
W304
Anode-cathode voltage too
low
Short circuit between
anode and cathode
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
W305
Anode-cathode voltage too
high
MSB defective
• Contact customer service
E306
Anode voltage incorrect
The anode voltage set
value is outside the
allowable range
• Perform a calibration
• Contact customer service
MSB defective
• Contact customer service
W310
Cathode 1 broken
Cathode defective
• Check that the other cathode has been
switched on
• Contact customer service
Connection to cathode
broken
• Contact customer service
VI-Board or MSB
defective
• Contact customer service