6.4 Measure
ü
The inlet flange on the upper side of the device is prepared for the pending
measurement. Your test specimen or the SL200 sniffer line is normally connected
there. See also "Connect recipient/test object to inlet flange [
47]".
ü
Alternatively, the sniffer line SL3000 is connected to the back of the device
(optional). In this case, the inlet flange on the top of the device should be closed
with a blank flange.
ü
Possible alternatives to the operating possibilities on the device are set up
(optional):
I/O module, bus module, remote control, wireless LAN connection to the notebook
or tablet. Also see "Accessories and spare parts [
detector via web browser (wireless LAN) [
139]".
1
Switch on the leak detector via the power supply switch.
2
Make sure that the desired operation mode is set (see "Select operation mode
3
Make sure that the correct basic settings and the settings for the current
measurement are carried out. See "Basic settings" and "Settings for the
measurements".
4
When performing a calibration, note the 20 minute warm-up time. See
75]".
5
In standby mode, initiate the measurement process by pressing the START
button.
ð
If you want to ensure that a current internal background is used in the
following measuring operation, please observe the notes in chapter “Change
73]“.
6
Keep track of the measurement results either as line, bar or circle chart, see
"Presentation of the measurement screen [
55]" and "Assembly of the
touchscreen".
7
To display the maximum leak rate (Q
max
) on the measurement screen during a
measurement, press the START button again. Also see "START button [
24] "
8
To measure small leak rates more easily, press the ZERO button when
required, see "Set and use the function ZERO [
9
Record the measured values if necessary, see "Switch on data recording and
81]" and "Design of touchscreen [
27]".
10
Switch the instrument off.
See also
2
Summary of Contents for UL6000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 125 152 ...
Page 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 ...
Page 151: ......
Page 152: ......