INFICON
Operation | 6
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
89 / 152
6.10.3 Exporting or importing sets of parameters
You are able to transfer saved sets of parameters from the internal memory to an
attached USB flash drive and to import the sets again.
ü
Supervisor
rights
1
> Parameter sets > Manage parameter sets
2
To export all sets of parameters to a USB flash drive (formatted in FAT32) if
needed, press
below the list.
3
To import all sets of parameters from a USB flash drive into the device, press
below the list.
ð
All sets of parameters on the USB flash drive will be saved in the device, if
the number of 10 is not exceeded. Using a PC you can delete redundant sets
of parameters on the USB flash drive before transferring.
ð
You must not modify parameters of a set of parameters on the USB flash
drive before importing. You cannot transfer these files to the device after the
modification.
6.11 Updating the software
The device has two different software versions: One for the user interface and one for
the basic unit. Each has its own independent versions number.
6.11.1 Update the user interface software
Import software updates using a USB flash drive.
Summary of Contents for UL6000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 125 152 ...
Page 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 ...